Reproducibility and accuracy of a wound measurement app
The wounds of 87 patients were measured using the Swift Wound app and a ruler. The skin surface temperature of 37 patients was also measured using an infrared FLIR™ camera integrated with the Swift Wound app and using the clinically accepted reference thermometer Exergen DermaTemp 1001. Accuracy measurements were determined by assessing differences in surface area measurements of 15 plastic wounds between a digital planimeter of known accuracy and the Swift Wound app. To evaluate the impact of training on the reproducibility of the Swift Wound app measurements, three novice raters with no wound care training, measured the length, width and area of 12 plastic model wounds using the app. High inter-rater reliabilities (ICC = 0.97–1.00) and high accuracies were obtained using the Swift Wound app across raters of different levels of training in wound care. The ruler method also yielded reliable wound measurements (ICC = 0.92–0.97), albeit lower than that of the Swift Wound app. Furthermore, there was no statistical difference between the temperature differences measured using the infrared camera and the clinically tested reference thermometer … read more